Software:MzXML2Search
From SPCTools
Revision as of 18:24, 24 May 2010 Jesse j (Talk | contribs) (→Running) ← Previous diff |
Revision as of 18:26, 24 May 2010 Jesse j (Talk | contribs) (→Running) Next diff → |
||
Line 13: | Line 13: | ||
-dta or -mgf or -pkl or -xdta or -odta or -ms2 output format (default dta) | -dta or -mgf or -pkl or -xdta or -odta or -ms2 output format (default dta) | ||
- | -F<num> where num is an int specifying the first scan | + | -F<num> where num is an int specifying the first scan |
- | -L<num> where num is an int specifying the last scan | + | -L<num> where num is an int specifying the last scan |
- | -C<n1>[-<n2>] "force charge(s)": where n1 is an integer | + | -C<n1>[-<n2>] "force charge(s)": where n1 is an integer |
- | specifying the precursor charge state (or possible | + | specifying the precursor charge state (or possible |
- | charge range from n1 to n2 inclusive) to use; this option | + | charge range from n1 to n2 inclusive) to use; this option |
- | forces input scans to be output with the user-specified | + | forces input scans to be output with the user-specified |
- | charge (or charge range) | + | charge (or charge range) |
- | -c<n1>[-<n2>] "suggest charge(s)": for scans which do not have a | + | -c<n1>[-<n2>] "suggest charge(s)": for scans which do not have a |
- | precursor charge (or charge range) already determined in the | + | precursor charge (or charge range) already determined in the |
- | input file, use the user-specified charge (or charge range) | + | input file, use the user-specified charge (or charge range) |
- | for those scans. Input scans which already have defined | + | for those scans. Input scans which already have defined |
- | charge (or charge range) are output with their original, | + | charge (or charge range) are output with their original, |
- | unchanged values. | + | unchanged values. |
- | -B<num> where num is a float specifying minimum MH+ mass, default=600.0 Da | + | -B<num> where num is a float specifying minimum MH+ mass, default=600.0 Da |
- | -T<num> where num is a float specifying maximum MH+ scan, default=4200.0 Da | + | -T<num> where num is a float specifying maximum MH+ scan, default=4200.0 Da |
- | -P<num> where num is an int specifying minimum peak count, default=5 | + | -P<num> where num is an int specifying minimum peak count, default=5 |
- | -I<num> where num is a float specifying minimum threshold for peak intensity, default=0.01 | + | -I<num> where num is a float specifying minimum threshold for peak intensity, default=0.01 |
- | -M<n1>[-<n2>]where n1 is an int specifying MS level to export (default=2) | + | -M<n1>[-<n2>]where n1 is an int specifying MS level to export (default=2) |
- | and n2 specifies an optional range of MS levels to export | + | and n2 specifies an optional range of MS levels to export |
- | -A<str> where str is the activation method, "CID" (default) or "ETD" | + | -A<str> where str is the activation method, "CID" (default) or "ETD" |
- | if activation method not in scans of mzXML file, this option is ignored | + | if activation method not in scans of mzXML file, this option is ignored |
- | -h use hydrogen mass for charge ion (default is proton mass) | + | -h use hydrogen mass for charge ion (default is proton mass) |
For example: | For example: |
Revision as of 18:26, 24 May 2010
Description
Converts mzXML files to SEQUEST dta, MASCOT mgf, and Micromass pkl files.
Getting the Software
Comes with the current version of TPP.
Running
MzXML2Search [options] *.mzXML
Options are:
-dta or -mgf or -pkl or -xdta or -odta or -ms2 output format (default dta) -F<num> where num is an int specifying the first scan -L<num> where num is an int specifying the last scan -C<n1>[-<n2>] "force charge(s)": where n1 is an integer
specifying the precursor charge state (or possible charge range from n1 to n2 inclusive) to use; this option forces input scans to be output with the user-specified charge (or charge range)
-c<n1>[-<n2>] "suggest charge(s)": for scans which do not have a
precursor charge (or charge range) already determined in the input file, use the user-specified charge (or charge range) for those scans. Input scans which already have defined charge (or charge range) are output with their original, unchanged values.
-B<num> where num is a float specifying minimum MH+ mass, default=600.0 Da -T<num> where num is a float specifying maximum MH+ scan, default=4200.0 Da -P<num> where num is an int specifying minimum peak count, default=5 -I<num> where num is a float specifying minimum threshold for peak intensity, default=0.01 -M<n1>[-<n2>]where n1 is an int specifying MS level to export (default=2)
and n2 specifies an optional range of MS levels to export
-A<str> where str is the activation method, "CID" (default) or "ETD"
if activation method not in scans of mzXML file, this option is ignored
-h use hydrogen mass for charge ion (default is proton mass)
For example:
MzXML2Search .mgf example.mzXML